Transmitter and Wiring Instructions

Scroll down the list to find your transmitter and wiring instructions.

Rosemount 54E & 5081 - With Pre-amp

Cable Colour CodingInstrument Terminal NameUniloc Terminal Number
PurpleTC3
BlueTC4
YellowTC5
RedReference7
RedReference8
WhitemV in10
Black- 5 VDC11
Green+ 5 VDC12
Note 1: The temperature compensator is 100 Ohm Pt RTD.
Note 2: Slide switch S1 toward the edge of the circuit board to select external amplifier.
Note 3: The preamplifier does not support diagnostic features, if any.

Rosemount 54E & 5081 - Without Pre-amp

Cable Colour CodingInstrument Terminal NameUniloc Terminal Number
BlackRTD Sense4
BlackRTD In5
RedRef In7
ClearmV In10
Note 1: Set Preamp location jumper to “Analyzer” in front panel
Note 2: A jumper is required between Terminal Numbers 7 & 8 on Terminal Block TB1.  This will disable any diagnostics for reference impedance in the analyser.  A second jumper is required between Terminal Numbers 3 & 4 on Terminal Block TB1.  This disables the automatic cable length compensation for the 100 Ohm Platinum Temperature Element (temperature must be calibrated manually in the transmitter, if required).  No manual calibration of the temperature element should be required for cable lengths of less than 16 feet.

Rosemount 1055 – with Preamp

For input Channel 1
Cable Colour CodingTerminal NameTerminal NumberTerminalBlock
PurpleRTD RETURN 16TB3
BlueRTD SENSE 15TB3
YellowRTD IN 14TB3
RedREFERENCE IN 16TB4
RedSOLUTION GND 14TB4
WhitepH / ORP IN 11TB4
Black- 5 V12TB3
Green+ 5 V11TB3
For input Channel 2 (if 1055 is a Dual Channel pH/ORP Analyser only)
Cable Colour CodingTerminal NameTerminal NumberTerminalBlock
PurpleRTD RETURN 27TB3
BlueRTD SENSE 28TB3
YellowRTD IN 29TB3
RedREFERENCE IN 26TB5
RedSOLUTION GND 24TB5
WhitepH / ORP IN 21TB5
Black- 5 V12TB3
Green+ 5 V11TB3

Rosemount 1055 – without Preamp

For input Channel 1
Cable Colour CodingTerminal NameTerminal NumberTerminalBlock
BlackRTD IN 14TB3
BlackRTD SENSE 15TB3
RedREFERENCE IN 16TB4
ClearpH / ORP IN 11TB4
Note 1: A jumper is required between Terminal Numbers 4 & 6 on Terminal Block TB4.  This will disable any diagnostics for reference impedance in the analyser.  A second jumper is required between Terminal Numbers 5 & 6 on Terminal Block TB3.  This disables the automatic cable length compensation for the 100 Ohm Platinum Temperature Element (temperature must be calibrated manually in the transmitter, if required).  No manual calibration of the temperature element should be required for cable lengths of less than 16 feet.
For input Channel 2 (if 1055 is a Dual Channel pH/ORP Analyser only)
Cable Colour CodingTerminal NameTerminal NumberTerminalBlock
BlackRTD IN 29TB3
BlackRTD SENSE 28TB3
RedREFERENCE IN 26TB5
ClearpH / ORP IN 21TB5
Note 1: A jumper is required between Terminal Numbers 4 & 6 on Terminal Block TB5.  This will disable any diagnostics for reference impedance in the analyser.  A second jumper is required between Terminal Numbers 7 & 8 on Terminal Block TB3.  This disables the automatic cable length compensation for the 100 Ohm Platinum Temperature Element (termperature must be calibrated manually in the transmitter, if required).  No manual calibration of the temperature element should be required for cable lengths of less than 16 feet.

PT91 - With Pre-amp

Cable Colour CodingInstrument Terminal NameUniloc Terminal Number
WhiteSignal1
RedReference2
Green+ V (5 V MAX)3
Black- V (5 V MAX)4
YellowTC (3K Ohm Balco)5
BlueTC (3K Ohm Balco)6